Insert dust covers into the laser fiber and optical temperature sensor connectors when
plugs not inserted (see orange arrows in Figure 7.32) and cover the end of the temperature
sensor fiber with the protective sleeve
7.4.
N
EURO
B
LATE
S
YSTEM
C
ONTROL
W
ORKSTATION
:
The Control Workstation is equipped with an Emergency Stop switch located on the front of the
monitor (see Figure 7.4) which intended to disable the system in an emergency. Check that the
E-Stop switch is released (red light should NOT be on). If it is, twist the red knob clockwise to
release.
•
Turn on the power switch at the side of the monitor (see Figure 7.4) and the monitor will
display NeuroBlate System logo launch screen.
•
Double-click on the
M*Vision for NeuroBlate System
icon to launch the software
application.
CAUTION:
The user must perform a
Shut down
under at the
Start
menu of the
Windows operating software before using the power switch to power down the
system. Failure to do so may result in software failure at restart.
CAUTION:
Full power shutdown of the system takes 2 minutes. After system
shutdown wait 3 minutes to restart power or the system may fail to start up.
Figure 7.4:
Control Workstation
Emergency-Stop
Power
Switch
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...