NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 106 -
Self-Test Step
Interface Platform Display
•
Perform the Laser aiming beam test.
•
The visible pilot laser provides a bright, red laser light
exiting the LDP tip in the correct orientation from the LDP
must be visible. Aiming the SideFire beam at a surgical
glove should produce a bright, red spot (see Figure
8.39).The FullFire should exhibit a bright red light
emanating from the effective focal area on the probe tip.
Figure 8.39: Laser Aiming Beam Test - SideFire®
Figure 8.40: Laser Beam Test - FullFire®
Note:
If the physician does not see a strong, visible red aiming laser beam exiting the SideFire
Probe (Figure 8.39) tip or emanating from fiber tip within the FullFire probe tip’s clear capsule
(Figure 8.40), ensure that LDP Laser Plug is fully inserted (two-clicks) into its connector on the IP
Connector Module. If the connected probe doesn’t produce a red laser beam, disconnect the
LDP and select a second LDP of the same size and configuration. Set the depth stop to the
correct setting and repeat the Laser Aiming Beam Test. If a failure occurs again, contact
Monteris service for diagnosis.
A bright red, visible laser spot should exit
the SideFire™ in same direction as the
protrusion on the Probe Locking Interface.
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...