NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 29 -
4.3.2.
Advanced Probe Driver, Robotic Probe Driver and LDP Warnings
The following are specific warnings that apply to the NeuroBlate Probe Drivers (APD
and RPD) and Laser Delivery Probe (LDP).
WARNING:
•
For single use only. Do not reuse, reprocess or re-sterilize. Reuse, reprocessing or
re-sterilization may compromise the structural integrity of the device and/or
lead to device failure which in turn may result in patient injury, illness or death.
Reuse, reprocessing or re-sterilization may also create a risk of contamination of
the device and/or cause patient infection or cross-infection, including, but not
limited to, the transmission of infectious disease(s) from one patient to another.
Contamination of the device may lead to injury, illness or death of the patient.
•
Ensure that all cables and umbilical in the vicinity of the MRI bore do not form
loops as this may result in heating and RF interference.
•
Contact between the cables and umbilicals with the inner surface of the MRI
bore may result in heating and RF interference.
•
The reduced diameter 2.2 series of LDP achieve target accuracy only if the
pathway is first created to the target with an appropriate dilator probe such as a
biopsy needle or stylet.
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...