NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 108 -
Using the following steps, the sterile LDP can now be advanced into the APD/RPD Follower and
into the brain until the LDP Locking Interface comes in full contact with the Mating Adapter on
the Follower (see Figure 8.6.31).
CAUTION:
Final insertion to PDP should always be done by hand, under full user
control where tactile feedback is important.
Figure 8.6.31: Laser Delivery Probe as Inserted into the APD/RPD Follower
Figure 8.6.32: Left: Correct LDP Insertion Technique; Right:
Incorrect
LDP Insertion Technique
•
Slowly insert the LDP through the Probe Driver Follower and visually align the LDP thumb
grip to the alignment mark (grey line) on the Mating Adapter (Figure 8.6.31)
•
Continue insertion until fully engaged to the Follower Mating Adapter and locked in
position; an audible click will be heard
WARNING:
To prevent inadvertent unlocking or slipping of the LDP Depth Stop
during insertion, hold the LDP at the bottom-most portion of the Locking Interface
while inserting into the brain as shown in Figure 8.6.32 to prevent potential injury.
Correct finger
grip location
Mating Adapter
Alignment Mark
LDP Thumb Grip
LDP Fully
Engaged to
Follower
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...