NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 78 -
Attachment Method 3 – IMRIS Side Rail Attachment
1.
Place both clamp pieces on each side rail by screwing clamp knobs in from bottom. Ensure
clamps are flush with end of table. If IMRIS clamp is in the way, slide further down rail as to
not interfere with the NeuroBlate attachment or procedure.
2.
Insert arms through side holes and tighten clamp knob onto the arm shaft. Note: Do not
over tighten the screw as this may damage the plastic.
g
h
Figure 8.43 g, h: Interface Platform to IMRIS Side Rail Arm Mounting
Attachment Method 4 – Siemens Table Slot Attachment
1.
Place both Table Blocks on the MRI table as shown.
Note:
There are two different arm
anchor types depending on table. The figures below highlight the differences.
2.
Slide the Table Anchors up through the second MRI Table Slot on each side. Insert
thumbscrews though the anchors and into the Table Blocks.
3.
Slide the IP Arm Tabs into the Table Block Slots (red arrows).
4.
Insert the Pivot Pins though the Table blocks and into the Interface Platform Arms (green
arrows).
Upper Clamp
Piece
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...