NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 128 -
Move slice plane cross hair until it MOST importantly move crosshair until the
intersects the yellow cross hair which Start position shown in GE Start dialog exactly
simply acts as a guide at this step matches the end point localizer position shown by
yellow text in top right view
•
Using the in plane rotation grab handles in the End Point (top right),
rotate the Graphical prescription until the projection as seen in the Start
Point (top left) view passes through the center of the default cursor
location.
4.
Acquire Parallel FSPGR Scan (1 sec)
5.
Prescribe NeuroBlate Thermal Sequence (or a “template” version of it having a single
slice and single phase to start)
o
Load the parallel (or stack) scan into all views. Use the “OK All” button when
loading series to ensure all views are populated with single parallel series
o
Perform an Erase All to allow for you to do a keyboard prescription
o
Setup View Options
•
Clear ALL or Erase All (mandatory)
•
Display Normal (mandatory)
•
Disable “Update ALL” (mandatory)
•
Enable “Cursors” or “Report Cursor” (mandatory) - in this case, ALL views
should now report the same cursor location.
•
Disable Ref Loc Lines (optional)
Click on rotation grab handle and
swing line while monitoring line
intersection with cursor in left view
Top Right View
“Endpoint cursor”
Top Left View
“Startpoint cursor”
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...