NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 125 -
8.8.2.1.
GE Thermal Sequence Prescription Procedure
•
While in the
Treat
task within M*Vision, select the
Scan Plane
icon from the task menu
on the right.
1.
Acquire Start Point Localizer (3-4 sec scan), using values from
Treat All Planes
o
Perform an Erase All to allow for you to do a keyboard prescription
o
Make sure that you have only a single image in each orientation (Spacing: 0.0)
Position entry dialog on v12.X and v15.X systems
o
Enter Center Position corresponding to
Scanning Range – Start
in All Planes View
o
Note: If position not allowed entry, Gradient Mode made need to be changed to
allow slice position outside optimal FOV (Grad Mode: Wide rather than Grad
Mode: Zoom )
2.
Acquire End Point Localizer (3-4 sec scan), using values from Treat-Treat All Planes
o
Perform an Erase All to allow for you to do a keyboard prescription
o
Make sure that you have only a single image in each orientation (Spacing: 0.0)
o
Enter Center Position corresponding to
Scanning Range – End
in All Planes View
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...