NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 126 -
Portion of Scan Plane Parameter Dialog showing information
for start and end localizer center position
3.
Prescribe Parallel FSPGR single image scan:
o
Copy/Paste/DoubleClick Parallel FSPGR
•
Make it active to begin prescription
•
Load the following series into MRI Console views (using the “OK” and
NOT the “OK ALL” button)
o
Start Point Localizer in top left view
o
End Point Localizer in top right view
•
Set the number of Images (or slices) to 1
•
Note: Both top views should show image 2 of 3 of the localizer which
is the sagittal view of each localizer in order to correctly display the
next step.
o
Setup View Options
•
Clear ALL or Erase All (mandatory)
•
Display Normal (mandatory)
•
Disable “Update ALL” (mandatory)
•
Enable “Cursors” or “Report Cursor” (mandatory) – this displays the
center yellow cross hair in view and reports its position in yellow text in
lower right of view.
•
Disable Ref Loc Lines (optional)
o
Check to make sure that the yellow text in lower right hand corner of each of the
two top views displays values that correspond to the values from the Treat-Treat
All Planes Dialog (those same values entered for the Start and Endpoint
localizers).
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...