NeuroBlate® System Instructions for Use
80469 Rev D – Nov 16, 2018
- 86 -
8.5.1.
Scan and Register Data in M*Vision using
Plan Register
Task
•
Ensure linear or rotary LDP motion is NOT active and acquire a 3-plane 2D localizing scan
•
Acquire a 3D Volumetric MRI scan with 1mm thick slices through head to include the entire
extent of the MRI Trajectory Wand within AXiiiS device
o
If not using AXiiiS, acquisition of this initial MR scan is required but its quality and
type is not relevant until the LDP is inserted later during the self-test procedure.
This initial scan is only needed to advance in the workflow and will be replaced when
this step is repeated. Once a probe is inserted, acquire a 3D Volumetric MRI scan
with 1mm thick slices through head to include the entire extent of the MR wand
attached to RPD (if applicable). Set this new scan as the
Master Series (see below)
.
•
Ensure the M*Vision software is set to the
Plan Register
workflow task
•
Push the acquired image data to the Monteris Control Workstation via the DICOM node set
up on the MRI system.
Once data is pushed from the MRI to NeuroBlate workstation, the data automatically loads into
the M*Vision software when in the
Plan Register
or
Treat Register
workflow task.
If multiple images are sent to the system, all data sets need to be registered to the Master Data
Set. Any of the image sets brought into
Plan Register
or
Treat Register
can be made the Master
series by right clicking on the image tab to the right and selecting
Set As Master
. Any data set
can be erased from the software by right clicking on the image and selecting
Clear
.
•
Co-register the acquired image data with previously loaded pre-planning image data (if any)
as described previously in section 8.1.2
Load and Register Image Data
.
o
If co-registration is required, visually inspect registration accuracy, and if needed,
use the manual correction described in 8.1.2
Load and Register Image Data
o
Select the
Verify Registration
to continue
8.5.2.
Create Region of Interest (ROI) Volumes within
Plan Volumes
Task
Using the M*Vision
Plan Volumes
workflow task, define treatment ROI(s) (if not already
defined) following the steps described in section 8.1.3
Create Region of Interest Volumes
.
Содержание Neuroblate
Страница 1: ...Instructions for Use ...
Страница 23: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 23 4 2 EQUIPMENT LABELS Dornier Laser ...
Страница 24: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 24 NeuroBlate System ...
Страница 152: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 152 ...
Страница 153: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 153 ...
Страница 154: ...NeuroBlate System Instructions for Use 80469 Rev D Nov 16 2018 154 ...