Chapter Six: Embedded Web-Based GUI and Diagnostics
Setup Mode
68
through Ethernet (see notes below), watch the MFC’s performance on the plot, and capture flow and other
parameter data from the MFC.
Note
For Analog units, sending a setpoint through Ethernet will not work unless the Analog checkbox
is selected. This checkbox tells the device to bypass the analog setpoint on the analog/power
interface. To begin sending setpoints through the analog interface again, either uncheck the
Analog checkbox or close the browser.
Note
If optional input control is enabled, set points (whether over analog I/O or Ethernet) will control
the optional input.
Figure 25: Embedded GUI, Plot Page in Setup Mode
Before a MFC leaves the factory the performance is checked with Nitrogen in a typical application. During
this test the control parameters are set so that the device’s performance is optimized. These parameters are
expected to work with our setup with the exception of the optional input control (analog 15 pin D only). If
optional input is enabled, you will see tuning PID parameters and a minimum command current offset that
can adjust for overshoot. To change the current PID variables in the device, enter in a value for the Kp
(gain), Ki (integral) and/or Kd (derivative) parameters and press the “Submit” button. For information on how
to tune your MFC for optimal performance please see
Tuning the MFC Mass Flow Controller
.
Содержание P9B
Страница 1: ...1042936 001 Rev A January 2013 MKS Type P9B Digital Mass Flow Controller Instruction Manual...
Страница 4: ......
Страница 10: ...List of References vi This page intentionally left blank...
Страница 20: ......
Страница 22: ...12...
Страница 23: ...13 MKS Instruments Inc MKS Calibration and Service Center...
Страница 24: ...14...
Страница 26: ...16 MKS Instruments Inc MKS MKS Calibration and Service Center...
Страница 27: ...17...
Страница 28: ...18...
Страница 32: ......
Страница 56: ...Chapter Four Analog and Digital Interface Digital Interface Input and Output Options 46...
Страница 64: ......
Страница 68: ...Chapter Five Ethernet Interface Setup and Configuration Digital Interface Input and Output Options 58...
Страница 86: ...Chapter Six Embedded Web Based GUI and Diagnostics Tuning the MFC for Optional PC 76...
Страница 92: ...Chapter Seven Maintenance Pressure Zero Adjustment 82 This page intentionally left blank...
Страница 99: ...Troubleshooting Chart Chapter Eight Troubleshooting 89 This page intentionally left blank...
Страница 100: ......
Страница 105: ...Appendix A Product Specifications Electrical Specifications 95...
Страница 111: ...Model Code Description 101...
Страница 112: ...Model Code Description Appendix C Health and Safety Form 102 Appendix C Health and Safety Form...