Chapter Four: Analog and Digital Interface
Analog Interface Input and Output Options
36
5. In selecting the appropriate type and wire size for cables, consider:
Voltage ratings.
Cumulative I
2
R heating of all the conductors (keep them safely cool).
IR drop of the conductors, so that adequate power or signal voltage gets to the device.
Capacitance and inductance of cables that handle fast signals (such as data lines or stepper motor
drive cables).
Some cables may need internal shielding from specific wires to others.
Analog Interface Input and Output Options
The P9B analog I/O MFC is available with either a 9 pin D male connector or a 15 pin D male
connector for providing power and signal I/O.
Table 8: Analog Interface – 9 Pin D Males Pinouts – Model Code A
Pin 1
Valve Open/Close: Apply +5 to+15 VDC to Open; Pull to ground or apply -5
to -15 VDC to Close.
Pin 2
Flow Output Signal, 0-5 VDC
Pin 3
+15 to +24 VDC Power Supply (+/-5% about nominal supply)
Pin 4
Power Common (See Note 2 below)
Pin 5
No Connection
Pin 6
Setpoint Input, 0-5 VDC
Pin 7
Signal Common
Pin 8
Signal Common
Pin 9
Valve Test Point
Notes
1. Chassis ground is not available on a separate pin. Instead, it is carried out through the cable
shielding. Be sure that the connector on the other end of the cable is properly grounded to its
chassis ground.
2. Signal common and power common are not connected within the device. The user must
connect the power common to the signal common at either the “tool-side” (recommended) or at
the MFC, but not at both locations. Connecting at both locations will cause ground loops.
3. The 0 to 5 VDC flow signal output comes from pin 2 and is referenced to pin 7 (signal
common).
4. Use any appropriate 0 to 5 VDC input signal of less than 1K ohm source impedance
referenced to pin 7 as the setpoint signal to pin 6.
Содержание P9B
Страница 1: ...1042936 001 Rev A January 2013 MKS Type P9B Digital Mass Flow Controller Instruction Manual...
Страница 4: ......
Страница 10: ...List of References vi This page intentionally left blank...
Страница 20: ......
Страница 22: ...12...
Страница 23: ...13 MKS Instruments Inc MKS Calibration and Service Center...
Страница 24: ...14...
Страница 26: ...16 MKS Instruments Inc MKS MKS Calibration and Service Center...
Страница 27: ...17...
Страница 28: ...18...
Страница 32: ......
Страница 56: ...Chapter Four Analog and Digital Interface Digital Interface Input and Output Options 46...
Страница 64: ......
Страница 68: ...Chapter Five Ethernet Interface Setup and Configuration Digital Interface Input and Output Options 58...
Страница 86: ...Chapter Six Embedded Web Based GUI and Diagnostics Tuning the MFC for Optional PC 76...
Страница 92: ...Chapter Seven Maintenance Pressure Zero Adjustment 82 This page intentionally left blank...
Страница 99: ...Troubleshooting Chart Chapter Eight Troubleshooting 89 This page intentionally left blank...
Страница 100: ......
Страница 105: ...Appendix A Product Specifications Electrical Specifications 95...
Страница 111: ...Model Code Description 101...
Страница 112: ...Model Code Description Appendix C Health and Safety Form 102 Appendix C Health and Safety Form...