Analog Interface Input and Output Options
Chapter Four: Analog and Digital Interface
39
In order to complete a true zeroing of the device, ensure the following conditions are satisfied prior to
beginning the procedure.
Device is installed in the orientation intended for final use (i.e. horizontal base down, vertical flow up,
etc.).
Device is powered at operating temperature, preferably for 30 or more minutes.
Devices subject to ambient temperatures other than room temperature (23º C) should be zeroed under
those conditions.
Pressure drop and flow across the device is reduced to zero. Depending on the gas panel configuration,
this may be done by one of the referenced procedures. See Chapter 7 “Maintenance” for zeroing
procedures.
How To Override the Valve (Controllers Only)
The valve override feature enables the control valve to be fully opened (purged) or closed independent of the
setpoint command signal.
If the MFC is equipped with a 9-pin Type D connector:
To Open
the valve, apply 5 to 15 VDC to Pin 1.
To Close
the valve, apply -5 to -15 VDC Low to Pin 1 or connect Pin 1 to Signal Ground.
Normal Setpoint operation occurs when Pin 1 is allowed to float.
If the MFC is equipped with a 15-pin Type D connector:
To Open
the valve by applying a +5 to 15 VDC to pin 4
or
connect pin 4 to the power ground pin.
To Close
the valve by applying a -5 to 15 VDC low to pin 3
or
connect pin 3 to the power ground pin.
Priority of the Valve Commands
The MFC executes commands based on a hierarchical command structure. The highest priority command is
Valve Close, followed by Valve Open, and then Setpoint Control. Therefore, if the flow controller is
operating under Setpoint Control, you can send a Valve Open command to force the valve to the full open
position.
Note
When both the Valve Close and Valve Open pins are pulled down, the Valve Close command
takes precedence and the valve is moved to the closed position.
Содержание P9B
Страница 1: ...1042936 001 Rev A January 2013 MKS Type P9B Digital Mass Flow Controller Instruction Manual...
Страница 4: ......
Страница 10: ...List of References vi This page intentionally left blank...
Страница 20: ......
Страница 22: ...12...
Страница 23: ...13 MKS Instruments Inc MKS Calibration and Service Center...
Страница 24: ...14...
Страница 26: ...16 MKS Instruments Inc MKS MKS Calibration and Service Center...
Страница 27: ...17...
Страница 28: ...18...
Страница 32: ......
Страница 56: ...Chapter Four Analog and Digital Interface Digital Interface Input and Output Options 46...
Страница 64: ......
Страница 68: ...Chapter Five Ethernet Interface Setup and Configuration Digital Interface Input and Output Options 58...
Страница 86: ...Chapter Six Embedded Web Based GUI and Diagnostics Tuning the MFC for Optional PC 76...
Страница 92: ...Chapter Seven Maintenance Pressure Zero Adjustment 82 This page intentionally left blank...
Страница 99: ...Troubleshooting Chart Chapter Eight Troubleshooting 89 This page intentionally left blank...
Страница 100: ......
Страница 105: ...Appendix A Product Specifications Electrical Specifications 95...
Страница 111: ...Model Code Description 101...
Страница 112: ...Model Code Description Appendix C Health and Safety Form 102 Appendix C Health and Safety Form...