How To Determine the Valve Configuration
Appendix D: MFC Sizing Guidelines
72
2. Select the valve configuration with the Cv value that is closest to, though larger than, the
Cv value calculated in step 1.
The Cv value represents the maximum flow rate for the unit. Choose the valve
configuration above your calculated Cv value to ensure that the unit can deliver the
required flow.
Valve Configuration Selection Guide
Valve Configuration
Nominal Range (N2) sccm
Cv x 10
5
1
10
2.44
2
20
4.88
3
50
12.21
4
100
24.42
5
200
48.84
6
500
122.11
7
1000
244.22
8
2000
488.44
9
5000
1221.11
10
10000
1924.47
11
20000
3848.94
12
30000
5773.41
Table 13: Valve Configuration Selection Guide