Chapter Four: Operation
How To Use the Optional Input (MFC only)
43
How To Use the Optional Input (MFC only)
The 1179 and 2179 units provide an optional input feature which allows them to control flow
based on 0 to 5 V signals from external sensing devices. A common application of this feature is
pressure control using inputs from a pressure transducer.
Implement the optional feature by simply routing the output from the desired external device to
the appropriate “optional input” position for the particular connector. Refer to Table 8, page 34,
and Table 7, page 33, for the pinout assignments. Voltage to the optional input overrides the
signal generated by the unit’s own internal flow sensor. The control electronics drives the valve
so that the optional input signal matches the set point. Provide the 0 to 5 V set point to the same
input pin as in standard flow control.
Metered flow output is still available on the standard output pin identified in the applicable pin
assignments. Refer to Table 8, page 34, and Table 7, page 33, for the pinout assignments.
Содержание Mass-Flo 1179A
Страница 3: ...116527 P1 Rev E 3 99 MKS Type 1179A and 2179A Mass Flo Controller and Type 179A Mass Flo Meter...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 54: ...How To Use the Optional Input MFC only Chapter Four Operation 44 This page intentionally left blank...
Страница 66: ...How To Adjust the Valve Preload MFC only Chapter Seven Troubleshooting 56 This page intentionally left blank...
Страница 70: ...Physical Specifications Appendix A Product Specifications 60 This page intentionally left blank...
Страница 74: ...Model Code Appendix B Model Code Explanation 64 This page intentionally left blank...
Страница 84: ...How To Determine the Valve Configuration Appendix D MFC Sizing Guidelines 74 This page intentionally left blank...