Chapter Five: Theory of Operation
General Information
45
Chapter Five: Theory of Operation
General Information
The 1179 Flow Controller measures the mass flow rate of a gas and controls the flow rate
according to a given set point. The control range is from 2 to 100% of Full Scale (F.S.) with an
accuracy of ± 1% of F.S.
Flow Path
Upon entering the flow controller, the gas stream passes first through the metering section of the
instrument for its mass flow to be measured. The gas moves on through the control valve for its
rate of flow to be regulated according to the given set point, and then exits the instrument at the
established rate of flow.
The metering section consists of one of the following:
•
A sensor tube for ranges < 10 sccm (N
2
equivalent)
•
A sensor tube and parallel bypass for ranges > 10 sccm (N
2
equivalent)
The geometry of the sensor tube, in conjunction with the specified full scale flow rate, ensures
fully developed laminar flow in the sensing region. The bypass elements, in those instruments
containing them, are specifically matched to the characteristics of the sensor tube to achieve a
laminar flow splitting ratio which remains constant throughout each range.
Measurement Technique
The flow measurement is based on differential heat transfer between temperature sensing heater
elements which are attached symmetrically to the sensor tube. This senses the thermal mass
movement which is converted to mass flow via the specific heat, C
p
, of the gas. The resulting
signal is amplified to provide a 0 to 5 VDC output which is proportional to mass flow.
Содержание Mass-Flo 1179A
Страница 3: ...116527 P1 Rev E 3 99 MKS Type 1179A and 2179A Mass Flo Controller and Type 179A Mass Flo Meter...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 54: ...How To Use the Optional Input MFC only Chapter Four Operation 44 This page intentionally left blank...
Страница 66: ...How To Adjust the Valve Preload MFC only Chapter Seven Troubleshooting 56 This page intentionally left blank...
Страница 70: ...Physical Specifications Appendix A Product Specifications 60 This page intentionally left blank...
Страница 74: ...Model Code Appendix B Model Code Explanation 64 This page intentionally left blank...
Страница 84: ...How To Determine the Valve Configuration Appendix D MFC Sizing Guidelines 74 This page intentionally left blank...