5-7
NOTE
z
Exercise care when pulling the front housing out. Be sure not to damage the cables
and connectors.
z
Avoid pressing the knob on the table.
2. Separate the front and rear housing and unplug the connecting cable between button
board and main board as well as the cable between the LCD and the main board.
Avoid pressing the
knob on the table.