7.4.1 How to Upgrade Software
Before software upgrade, check software version by selecting [
Main Menu
]
→
[
Maintenance>>
]
→
[
Factory Maintenance>>
]
→
enter the required password
→
[
Software
Version>>
]. If your BIOS program has a version prior to 2.0, upgrade it to version 2.0 or
greater before upgrading other programs. If you use System Update Tool (PN:
110-001595-00) version 2.0 or greater, make sure that BIOS program version is 2.0 or greater
before upgrade. Please refer to the following procedure to upgrade the software:
1. Connect the patient monitor to a PC through a crossover network cable. Set the IP to
77.77.1.XX, the subnet mask to 255.255.255.0.
2. Run the System Update Tool in the PC. Select [BeneViewT5/T6/T8] in the [Machine
Type Selection] window and confirm.
3. Select [Select Package] in the main screen of the Mindray Patient Monitor Software
Upgrade Tool. Choose the package you want to upgrade and click [Start].
4. Start up the patient monitor and press quickly the [S[Main Menu] hardkeys to
enter upgrade mode. Determine if the software is upgraded successfully according to the
prompt message of the patient monitor when finishing upgrading the software.
If your BIOS program has a version prior to 4.2.0 (FDA), press [Silence
]
hardkey before you
start the patient monitor, and do not release it until you enter the upgrade screen.
Software upgrade is taken in the sequence of BIOS program
→
system software package
(including system software, language library, and BMP resource files)
→
FPGA package
(including mother board FPGA software and audio data files)
→
module rack FPGA package
→
MPM module software.
Содержание BeneView T5
Страница 1: ...DPM 6 Patient Monitor Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 16: ...1 4 FOR YOUR NOTES ...
Страница 32: ...2 16 FOR YOUR NOTES ...
Страница 68: ...3 36 FOR YOUR NOTES ...
Страница 80: ...4 12 FOR YOUR NOTES ...
Страница 118: ...5 38 4 Remove the module infrared communication backboard and disconnect all the cables from it ...
Страница 124: ...5 44 5 Unscrew the two M3 8 screws on the parameter board Then pull off the parameter board rightward as shown below ...
Страница 126: ...5 46 FOR YOUR NOTES ...
Страница 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Страница 163: ...6 37 6 6 7 BIS Module Exploded View ...
Страница 180: ...6 54 FOR YOUR NOTES ...
Страница 204: ...A 14 FOR YOUR NOTES ...
Страница 205: ......