3-12
4. Open the relief valve and vent a certain standard gas or gas mixture. Then adjust the
relief valve and make sure the flowmeter reading is stable and within 10 and 50 L/min.
5. In
the
[
Calibrate AG
] menu, the concentration and flowrate of each measured gas are
displayed.
If the difference between the measured gas concentration and the actual one is t
very small, a calibration is not needed.
If the difference is great, a calibration should be performed. Select [
Calibrate >>
]
to enter the calibrate menu.
6. Enter the vented gas concentration. If you use only one gas for calibration, set other
gases’ concentration to 0.
7. Select
[
Start
] to start calibration.
8. If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. If the calibration failed, the message [
Calibration Failed!
] is displayed. In
this case, perform another calibration.
CAUTION
z
Calibrate the O
2
module, If it has been transported for long distance.
Содержание BeneView T5
Страница 1: ...DPM 6 Patient Monitor Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 16: ...1 4 FOR YOUR NOTES ...
Страница 32: ...2 16 FOR YOUR NOTES ...
Страница 68: ...3 36 FOR YOUR NOTES ...
Страница 80: ...4 12 FOR YOUR NOTES ...
Страница 118: ...5 38 4 Remove the module infrared communication backboard and disconnect all the cables from it ...
Страница 124: ...5 44 5 Unscrew the two M3 8 screws on the parameter board Then pull off the parameter board rightward as shown below ...
Страница 126: ...5 46 FOR YOUR NOTES ...
Страница 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Страница 163: ...6 37 6 6 7 BIS Module Exploded View ...
Страница 180: ...6 54 FOR YOUR NOTES ...
Страница 204: ...A 14 FOR YOUR NOTES ...
Страница 205: ......