5-25
5.3.18 Removing the Main Support
1. Unscrew the five M3
×
6 screws and disassemble the main support.
Note
z
Be sure to remove the power module first before removing the main support.
5.3.19 Removing the Interface Board Assembly
1. Release the clip that locks the interface board and push it outwards.