5-29
3. From the left side, remove the cable that connects the SMR interface board and the SMR
communication board. Then take out the SMR from its housing.
4. Release the two clips and take out the SMR interface board. Be sure not to damage the
snap slot on the left side.
5. Remove the LED indicator, the light conductor and the cable that connects the SMR
communication board and the LED board.
Housing
SMR