5-33
2. Unlock the snap close by pressing it down about 1 mm with a flat screwdriver. At same
time, push the snap plate forward with your thumb until the snap close separate from its
counterpart. Lift the snap plate with the flat screwdriver and remove it from the ICG
module. In case the snap close relocks the snap plate when the end of the snap plate
reaches the snap close, push the snap plate forward softly with the flat screwdriver.
3. Unscrew
the
M3
×
6 screw. Then, press down, in turn, the two clips that engage the front
panel. At same time, separate the front panel from the module’s outer housing.
Snap plate