3-34
Block the gas inlet of the module or watertrap. The sidestream
CO
2
flowrate is slower than 10ml/min and an alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
Yes No
The displayed CO
2
value is within 6
±
0.05
%
.
Yes No
Miscrostream CO
2
test and calibration
Block the gas inlet of the module or watertrap. An alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
Yes No
The displayed CO
2
value is within 6.0
±
0.3
%
Yes No
AG test and calibration
When AG flowrate is slower than 10ml/min, an alarm of AG
Airway Occluded is given. It indicates that there is no leakage.
Yes No
The fan inside the AG module works properly.
Yes No
The measurement accuracy of CO
2
, N
2
O, O
2
and AA (AA
represents an anaesthetic agent) meets the product specifications
in the Operator’s Manual.
Yes No
ICG test
The measured results are as follows: HR=60±2 bpm,
VI=61±4/1000s, TFC=32±2/kOhms.
Yes No
The measured results are as follows: HR=70±2 bpm, VI=48±4
/1000s, TFC=32±2 / kOhms.
Yes No
BIS test (you can select either method to perform the test)
Method 1: The BIS value measured on healthy, wide-awake
adult is within 80-100.
Yes No
Method 2: Connect to the BIS simulator to perform a cyclic
impedance check. The EEG wave and BIS numeric are displayed
on the monitor.
Yes No
RM test
The displayed TV is within 500±50ml and RR is within
20±1rpm.
Yes No
CCO/SvO
2
test
The CCO/SvO
2
numerics displayed on the patient monitor and
Vigilance monitor are consistent.
Yes No
The waves (at the ECG signal output port) displayed on the
oscillograph are consistent with the ECG calibration waves
displayed on the monitor screen.
Yes No
The amplitude of electrical level at the signal output port of
MAP, CVP and SpO
2
are 5±0.25V, 5±0.25V and 10±0.5V
Yes No
Содержание BeneView T5
Страница 1: ...DPM 6 Patient Monitor Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 16: ...1 4 FOR YOUR NOTES ...
Страница 32: ...2 16 FOR YOUR NOTES ...
Страница 68: ...3 36 FOR YOUR NOTES ...
Страница 80: ...4 12 FOR YOUR NOTES ...
Страница 118: ...5 38 4 Remove the module infrared communication backboard and disconnect all the cables from it ...
Страница 124: ...5 44 5 Unscrew the two M3 8 screws on the parameter board Then pull off the parameter board rightward as shown below ...
Страница 126: ...5 46 FOR YOUR NOTES ...
Страница 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Страница 163: ...6 37 6 6 7 BIS Module Exploded View ...
Страница 180: ...6 54 FOR YOUR NOTES ...
Страница 204: ...A 14 FOR YOUR NOTES ...
Страница 205: ......