5-24
3. Lift the power board with a small flat screwdriver. Then, turn it over.
4. Unplug the cable between the AC input filter and the power board, then remove the
power board.
NOTE
z
Since the power board may be adhered to the insulator, be careful not to damage
the parts, connectors and cables on the power board when prizing it.
Equipotential Pillar