5-43
3. Disconnect the cable of the infrared communication board, and release the snap lock to
remove the infrared communication board.
4. Disconnect the cables of the NIBP inflation pump and the relief valves from the
parameter board. Snip off the cable ties to remove the pump. Release the snap locks to
remove the valves.
Snap