5.5.3 Disassembling the New MPM Module
1. Remove the contact screws and the spanner by referring to steps 1 to 3 as described in
5.5.1 Disassembling the ICG Module.
2. Unscrew the two M3×6 screws. Then press down, in turn, the four clips that engage the
front cover with a small flat-bladed screwdriver. At the same time, pull off the rear
cover.