Safety Features
11.5
Safety Switches on the Microscope
When the safety switches are triggered, the light path of the laser beam is interrupted.
Figure 47:
Safety switches (1) and (2) on the inverted microscope (left); Safety switches (3) on the mirror
housing for the upright microscope (right)
Position of
the safety
switch
Is triggered
by
Is triggered
if
Function
1
Transmitted-light
illuminator arm
The illuminator arm is
tilted (e.g. for working
on the specimen)
Prevents laser light while
working on the specimen
2
Motorized
changeover from
eyepiece mode to
scan mode
The path-folding
mirror for the scan
head is swung out by
motor
Prevents stray light in the
eyepiece if the user switches
from eyepiece observation to
confocal observation
3
Pushbutton in the
port apertures of the
mirror housing
One of the two
apertures on the
mirror housing is free
(no external detector,
no halogen lamps, no
cover)
Prevents the escape of laser
light if the user removes
components from the mirror
housing.
11.6
Warning Messages
•
Warning messages in the event of a defective shutter:
The shutters are monitored for
safety. If one of the shutters is defective or not functioning properly, a warning message
is shown on the monitor within the LAS AF. For safety reasons, the shutters remain
closed. No further use of the laser is possible. In this situation, the system must not be
operated. Contact the Leica branch office in your country or your contact person.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......