112
LAS AF
• Step
Measurements
: Here, you can define multiple FCS measuring points in the
previously acquired xy or xz image or xyz or xzy stack, then carry out consecutive and
automatic FCS measurements defined in the
Setup FCS
step. The list of measuring points
can be saved and reloaded.
In each step, different tabs are shown in the tab area:
• On the
"Workflow"
tab you will find a suggested workflow and the functions associated
with it.
• On the
"Experiments"
tab, the image data are stored.
• On the
Acquisition
tab, you can configure the settings for image acquisition, such as
scan speed, zoom or pinhole diameter, or you can define a series of FCS measurements.
In all steps of the Wizard, you can access continuous scanning or acquire a single image or
z or y-stack at any time with "
Live
",
Capture image
and
Start
. The conditions for image
acquisition that you defined in
Setup Imaging
are used for this.
Then, if you carry out the FCS measurement in step
Setup FCS
and do not move the
crosshair, the scanner is simply parked in the center of the image.
After the completion of the FCS measurements, you can close the Wizard by selecting the
operating mode
TCS SP8
(
Figure 91:
Closing the Wizard
FCS Measurements in Solutions
Even when carrying out FCS measurements in solutions, you have to
acquire an xy image before the first FCS measurement. This is necessary
so that the scanner can become oriented in the scan field. The
information content of this image is not relevant here.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......