201
F(L)CS Data Acquisition
Figure 178:
Loading and saving FCS-specific instrument parameter settings
18.4
Measurements Step – FCS Measurement Time Series at
Multiple Points
After optimizing the FCS measurement conditions, it is possible to have FCS measurements
run automatically at previously defined measuring points with the previously defined
settings. To do so, go to the
Measurements
step.
18.4.1
FCS Network Connection
The LAS AF and SymPhoTime programs are synchronized via a network connection. Each
FCS measurement started in LAS AF in the
Measurements
step of the FCS Wizard generates
a new data file in SymPhoTime. In addition to the synchronization, the network also enables
transmission of relevant information such as the file name, instrument parameter settings
(IPS), or comments entered by the user in LAS AF.
During the FCS measurement, an online correlation curve is displayed that is automatically
saved along with the raw data in SymPhoTime. After the data acquisition, a more detailed
data analysis can then be carried out offline.
Do not change Acquisition mode any more
Do not make any more changes to the
Acquisition mode
used to acquire
the last image.
Observe the SymPhoTime manual
For more detailed information about data analysis, refer to the
SymPhoTime manual.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......