149
FLIM Data Acquisition
5. If you want to view an online lifetime histogram, close the fast FLIM image in
SymPhoTime. Click the symbol
Show Oscilloscope...
and select the
TCSPC histogram
tab.
6. To cancel specimen illumination, select the
Stop FLIM Test
button in the LAS AF
software. This will also stop the online FLIM display in SymPhoTime.
The instrument parameters set for the FLIM measurements (laser lines and associated
intensity, pinhole, scan mode, image size etc.) are kept as separate FLIM settings that are
independent from the image acquisition setting defined in the
Setup Imaging
step. The FLIM
settings defined in the
Setup FLIM
step are automatically transferred to the
Measurements
step.
16.2.9
Count Rate Monitor
The count rate of each channel is displayed in the
Setup
tab. For FLIM, the count rate should
be between 10,000 and 1,000,000 cps (counts per second).
Figure 135:
Count rate monitor in the FLIM Wizard
In addition, the count rate monitor shows the number of photons in the brightest pixel of the
image. This value allows you to evaluate the image quality. The higher the number of
fluorescence photons (counts) per pixel, the more accurately the lifetime can be
determined. This value is updated regularly whenever a new scanning operation starts in
the
Setup FLIM
step of the FLIM Wizard.
16.2.10
Loading and Saving FLIM-specific Instrument Parameter Settings
The instrument parameter setting (IPS) for FLIM can be saved and loaded in the same way
as IPSs for normal intensity imaging.
Save the FLIM settings in the
Setup FLIM
step and settings for intensity image acquisition in
the
Setup Imaging
step.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......