Switching On the System
13
Switching On the System
The TCS SP8 SMD is available in different versions. You must precisely follow the switch-on
sequence that applies to your system variant. During FLIM experiments, it is also possible
to operate the system with a compact supply unit.
13.1
Confocal System with Flexible Supply Unit
NOTICE
Damage is possible when contacting the specimen stage
When using an inverted microscope, the illuminator arm must be swung
back before the system start and LAS AF start because the stage can
damage the condenser, the objective or the specimen during
initialization.
1. If you are using a fluorescence lamp EL6000, switch it on first.
Figure 61:
EL6000
2. If you are using a HyD RLD, switch it on at its supply unit (see
). You can find
additional information on the HyD RLD in
.
3. Switch on the workstation and the microscope on the main switch board of the flexible
). You do not have to start the operating system—it
starts automatically when you switch on the workstation. Wait until the startup process
has concluded.
NOTICE
Objective damage is possible when contacting the specimen stage
When using an upright microscope, the specimen stage must be moved
down before the system start and LAS AF start because it could come
into contact with the objective during initialization and damage the
objective.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......