199
F(L)CS Data Acquisition
Figure 176:
Starting the FCS test and applying the measuring point
5. To end illumination of the specimen, click the
Stop FCS Test
button in LAS AF. This will
also stop the online correlation display in SymPhoTime.
6. If you would like to reuse a tested FCS measuring point later for the 'correct' FCS
measurement, click
Add Point
) in the
Setup
tab. This measuring point
is then written to a position list, which is used in the
Measurements
step in LAS AF. Do
not make any more changes to the zoom. Changing the zoom in the
Setup FCS
step would
render the position list invalid.
The instrument parameters set for the FCS measurements (laser lines and associated
intensity, pinhole, AOBS) are kept as separate FCS settings that are independent from the
image acquisition setting defined in the
Setup Imaging
step. The FCS settings defined in the
Setup FCS
step are automatically transferred to the
Measurements
step.
18.3.6
Count Rate Monitor
The count rate of each channel is displayed in the
Setup
tab. For FCS, the count rate should
be between 10,000 and 500,000 cps (counts per second).
If the two detection channels are reversed, change the allocation in SymPhoTime. The
allocation of the channels is preset and should not be changed.
• FCS Channel A: Check box
1
• FCS Channel B: Check box
2
Saving the data in Measurements
This curve is not saved. If you want to save your data, you have to work in
the
Measurements
step.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......