177
F(L)CS Data Acquisition
18
F(L)CS Data Acquisition
With the TCS SP8 SMD, you can carry out point measurements and FCS and FLCS data
acquisitions are possible.
With sensitive detectors such as APDs, FCS measurements can be carried out. In all point
measurements, the fluorescence is recorded at a certain position in the specimen and
analyzed in the SymPhoTime software.
If no pulsed lasers are attached to the system or if lifetime information is not needed,
continuous wave lasers from TCS SP8 can be used for the fluorescence excitation.
18.1
Preparing the FCS Measurement
This chapter includes steps that are helpful in preparing for an FCS measurement, not the
description of the FCS measurement itself. That information can be found in
to
18.1.1
Selecting an Objective
• Select a suitable objective: Since with FCS variations the focus size and shape have a
particularly significant impact, we recommend using water immersion objectives for in
vitro assays and in vivo experiments.
• For good FCS results, adjust the cover slip thickness with the correction ring (if available
with the objective used).
• Select an objective with a high numerical aperture in order to maintain the smallest
possible observation volume and yet to collect as many emitting photons as possible.
We recommend the water immersion objective
HCX PL APO 63x/1.2W CORR CS
0.14 to 0.18
with manual or motorized correction ring. This objective functions
exclusively with specimen slides or (glass-bottomed) measurement chambers that are
140
μ
m to 180
μ
m thick (e.g. cover slip types #1.0 or #1.5).
18.1.2
Calibrating the Positioning Accuracy of the FCS Measuring Point
The measuring point is positioned via the scanner. Positioning accuracy can be optimized
using a calibration function. Repeat this procedure about every 3 days. If you determine that
the system is stable, you can also increase the time between fine-tuning sessions.
Use a bleach specimen with fixed fluorescence that is capable of being excited with a
wavelength in the visible range and can also be photobleached with higher laser intensity.
The bleach specimen may be a chroma slide (fluorescent dye that has been allowed to dry
onto a cover slip), or you can mark the top side of a cover slip with a fluorescent marker.
1. Switch the confocal system on and start the LAS AF software as described in the
instructions in
. Select the conventional mode.
2. Activate the lasers in the
Configuration
step,
Lasers
button. Set the
Gain
for the argon
laser to a value between 20 and 30 %.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......