101
Switching On the System
power switch is in the "ON" position.
3. Start the laser driver and switch the laser on. If you are using the PDL 828 Sepia II laser
driver, set the key switch to the "ON" position. Prior to data acquisition, the laser head
should be operated for a few minutes to stabilize. If you use a PDL 828 Sepia II, start the
specific Sepia software on the SMD workstation and set the laser settings there (refer
to corresponding manuals).
4. Start the SymPhoTime software on the SMD workstation.
5. Open an existing workspace by selecting
Load Workspace
in the
File
menu or create a
new workspace by selecting
New Workspace
in the
File
menu (see SymPhoTime
manual).
6. Set the correct fluorescence filter on the corresponding filter holder position of the
external PicoQuant detector unit or on the Leica TCS SP8. Select the correct SMD filter
cube in the X1 port adapter (
). If you have a dual-channel APD
detector unit, set the variable beam splitter unit to the correct position (see dual channel
detector unit manual for details).
7. Be sure that no light falls on the detectors. Either switch on the power supply or, if using
a DSN 102 power supply, start the respective detector by selecting the corresponding
channel button on the DSN 102.
The SMD upgrade is now ready to use. The communication between the SP8 confocal
system itself and the SMD upgrade electronics starts at the beginning of the data
acquisition.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......