108
LAS AF
Figure 87:
Overview of the FLIM Wizard steps
• "
Setup Corr-ring
" step Some immersion objectives have a correction ring to optimally
correct for differences in the refractive index between immersion and specimen and
temperature effects. An optimum adjustment of the correction ring correlates with the
best possible optical resolution of the system. In the
Setup Corr-ring
step, the beam path
is preconfigured for a fast and efficient adjustment of the correction ring.
•
Setup Imaging
workstep: Here, you can establish and optimize the conditions for the
acquisition of a reference image. In the
Setup Imaging
step, the FLIM data acquisition
cannot be started.
• Step
Setup FLIM
: Here, you can define and optimize instrument parameters for the FLIM
measurement. The FLIM measurement instrument parameters (such as selection of the
laser line and intensity, image size, pinhole size, etc.) are managed independently of the
parameters for image acquisition in the
Setup Imaging
step. The optimal laser intensity
can be determined with a FLIM test measurement (selecting the
FLIM Test
button).
• Step
Measurements
: Here, you can define the FLIM experiment (exposure time of the
individual FLIM images, definition of FLIM volume stacks, FLIM time series or, for internal
SP FLIM, FLIM lambda stacks).
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......