181
F(L)CS Data Acquisition
18.1.3
Testing the Positioning Accuracy
In order to test the positioning accuracy of the scanner on different pixels, the multipoint
functionality for FCS can be used for a different purpose.
This test is not necessary to ensure the functional capability of the system as such; you use
it only to confirm the performance capability of the system for yourself.
To carry out this test, proceed as follows:
1. Switch off all external detectors (the APD detector unit or the APDs with the DSN)
electrically.
2. Start the SMD workstation.
3. Start the SymPhoTime software.
4. Create a new working area by selecting
New Workspace
in the
File
menu.
5. Using your bleach specimen, create a well-illuminated xy image as described in
in the
Setup Imaging
step of the FCS Wizard.
Use the photomultipliers for this (
).
Use the zoom that you want to test.
Use the same scan speed as the one you used previously when calibrating the scanner.
6. Select the
Measurements
step in the FCS Wizard (
Observe the SymPhoTime manual
You will find information about creating a new working area in the
SymPhoTime manual.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......