FLIM Data Acquisition
The following window is displayed:
Figure 130:
Oscilloscope window in SymPhoTime
4. In the
Acquisition
field, you can change the resolution under
resol.
Please confirm the
settings by selecting
apply
. To cover the whole range between two consecutive laser
pulses at the maximum resolution possible, the following settings are recommended:
Table 20: Recommended resolution for data acquisition with different laser frequencies
The default is a setting of 16 ps suitable for a laser repetition rate down to 20 MHz.
16.2.5
2-Laser PIE (405, 470, 640 nm)
PIE (pulsed interleaved excitation) means that two lasers pulse alternatingly. This method is
used to identify crosstalk and cross excitation and remove them from the signal. In the SMD
system, this option is available for certain combinations of pulsed diode lasers (405+470,
405+640, 470+640). Exactly two diode lasers have to be selected in order to activate 2-laser
PIE. If this is not the case, the corresponding radio button is grayed out.
Laser frequency
Resolution of data acquisition
80 MHz (MP laser, white light laser)
4 ps
40 MHz (pulsed diode laser, white light laser)
8 ps
20 MHz (pulsed diode laser, white light laser)
16 ps
10 MHz (pulsed diode laser, white light laser)
32 ps
5 MHz (pulsed diode laser)
64 ps
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......