133
FLIM Data Acquisition
16.1.2.5
Using White Light Lasers
1. Select the desired laser lines in the
Whitelight
dialog in
Beam Path Settings
.
2. Make sure that the AOBS is at fluorescence.
3. Set the desired intensity using the AOTF slider.
4. If the system has a multifunction port (
MFP
), it should be set to
Substrate
.
5. Open the
Acquisition
tab and adjust the rotation of the scan field to 0° using the slider
).
Figure 118:
Adjusting the diameter of the pinhole and rotation of the scan field
16.1.3
Adjusting the Pinhole for Image Acquisition
For VIS and UV excitation, set the
Pinhole
diameter to
1 Airy
; for MP excitation set it to the
maximum size (
).
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......