155
FLIM Data Acquisition
5. Define the duration of the individual FLIM images as described in
. For
recording a FLIM time series, use either
Duration of each image
or
Repetitions
, because
the last option, (
Acquire until max ___ photons/pixel
), does not allow a fixed period of
time to be specified.
6. You can define a time series using the time series dialog (
). The following
inputs are possible:
•
Time Interval
: Defines the time between the beginning of one FLIM image and the
next image. This duration cannot be shorter than the duration entered for a single
FLIM image. To specify the smallest possible duration, select
Minimize
.
•
Duration
: Here, enter how long the entire time series is to last. The corresponding
number of FLIM images is acquired. The duration entered is automatically corrected
so that all FLIM images can be completed.
•
Frames
: You can specify the number of FLIM images to be acquired. The maximum
specified number is 1000.
Figure 140:
Defining an FLIM time series
7. Enter the desired value in the time series dialog.
8. Start the FLIM time series acquisition with the image acquisition button
Run FLIM
.
In LAS AF, you obtain a time series of averaged images and, in SymPhoTime, a series of
FLIM images.
16.3.6
Defining a Time Series of xyz or xzy FLIM Stacks
Proceed as follows to take a time series of FLIM stacks:
1. Select the specimen position.
2. Select either
Acquisition Mode xzyt
or
xyzt
as the scan mode on the
Acquisition
tab.
Optimize the instrument parameter setting for FLIM measurement.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......