52
Operating Manual
OMBG3000 – Iss.01
© Copyright 2016
Read-Only 16-Bit Registers
Register
Address
(Dec) (Hex) Parameter
Content Information
0
0000h Run status
0 – running process
1 – running with non-critical fault
2 – stopped by user (outputs frozen)
3 – stopped by user (fixed at safe
values)
4 – stopped with critical fault
(outputs frozen)
32
0020h Time and date: year e.g. 2016
33
0021h Time and date:
Month
1-12
34
0022h Time and date: day
1-31
35
0023h Time and date: hour 0-23
36
0024h Time and date:
minute
0-59
37
0025h Time and date:
second
0-59
48
0030h Current sample
point
1-4
128
0080h Service due: year
e.g. 2016
129
0081h Service due: month 1-12
130
0082h Service due: day
1-31
512
0200h Alarm 1 Status
0 – No alarm
1 – Triggered
2 – Triggered and in recovering zone
4 – Latched
8 – Muted
9 – Triggered alarm and muted
10 – Recovering alarm and muted
12 – Latched and muted
513
to
519
0201h
to
0207h
Alarm 2 to alarm 7
Status
Repeated as per alarm 1
768
0300h
Sample point 1 last
reading: year
e.g. 2016