Operating Manual
147
© Copyright 2016
OMBG3000 – Rev1
7.5
Replacing the Inline PTFE Filter
Warning
Do not open when an explosive atmosphere is present.
Mains voltages are present within the BIOGAS 3000
system and great care needs to be observed. If there is
any uncertainty, seek advice from a professional or
isolate the supply. Alternatively, contact our technical
support team at (800) 968-2026 or email
(Please note a charge may be applicable.)
The inline PTFE filter should be replaced if showing signs of contamination or
saturated with liquid. Failure to replace the filter will result in the gas flow
being restricted or blocked to the BIOGAS 3000 module. In addition, it will
cause a flow fail error.
Op. Image
Instruction
1
From the ‘Gas Readings
Screen’, press the middle soft
key to ‘Stop Sampling’.
Note: If required, press
the middle soft key once
more to fix the outputs at
their safe values.
2
Ensure the drain, gas inlet, and
gas outlet valves are closed.