Operating Manual
49
© Copyright 2016
OMBG3000 – Rev1
Wire the outputs in accordance with
. For optimum
performance, it is recommended that screened twisted pair cable be used.
Note: For cable conductor sizes and cable insulation requirements, please
refer to section
Cable Conductor Sizes and Insulation Requirements
When wiring the outputs the twisted pairs must be as follows:
Terminal Colour
Wiring Information
Pair
Orange
Signal A (RS485) ‘+’
Pair one
Blue
Signal B (RS485) ‘-‘
White
Logic Ground (0V) ‘LG’
Pair two
Wiring Diagram 2 – Modbus
adequately within the protective sheath of the ferrule.
Terminals that are wired should be tightened to a
minimum of 0.6N
m. Failure to tighten to this
requirement could make the equipment unsafe
resulting in a hazard and invalidate the ATEX/IECEx
certification.
The input voltage range to the BIOGAS 3000 Modbus
terminals must not exceed -7 to +12V and the current
must not exceed 83.3mA. Operating outside of this
range will invalidate the ATEX/IECEx certification.