Operating Manual
General Operating Instructions
115
© Copyright 2016
OMBG3000 – Rev1
Screen 30 - Fit New External Cell confirmation
2)
Press the right soft key to ‘Exit’ and return to the ‘Calibration’ menu.
3)
If fitting a new cell or complete module, confirm the cell type and range,
and press the left soft key to ‘Continue’.
Screen 31 - Input new Slope and Offset
Note: Alongside the new cell or complete module there will be a
calibration certificate containing the new slope and offset figures. Both of
these values need entering in to the appropriate fields in the above
screen.