168
Operating Manual
OMBG3000 – Iss.01
© Copyright 2016
9.6
Analogue Outputs Not Working
Please refer to section
in this operating manual to ensure the
outputs have been wired correctly. If the wiring method is not suitable for the
input systems configuration, please contact your our technical support team at
(800) 968-2026 or email
If the system is wired correctly, ensure the analogue outputs have been
configured correctly; refer to
Check that your 4-20mA input channel is being interpreted correctly; refer to
Finally, check fuse 5, which is used to provide power to the 4-20mA loop.
If the problem persists, please contact our technical support team at (800) 968-
2026 or email
9.7
Modbus Outputs Not Working
Please refer to section
ensure that the configuration of the Modbus port is correct and that the wiring
is correct as per
Note: If the BIOGAS 3000 is last on the bus, ensure the termination is set
to ‘On’. Likewise, if it is not, ensure the termination is set to ‘Off’.
In addition, also check that no two nodes on the bus have the same slave
address. If this is the case, the BIOGAS 3000 slave address can be changed by
referring to section
If the problem persists, please contact our technical support team at (800) 968-
2026 or email
9.8
Low Flow / Flow Fail
If a flow fail occurs, this is classified as a critical fault and sampling will be
stopped.