Operating Manual
137
© Copyright 2016
OMBG3000 – Rev1
Inspect, and replace if required, the coalescing catchpot filter (see
Inspect, and replace if required, the inline PTFE filter (see
Perform a pressure test to ensure there are no leaks (see
Perform a gas check to determine the accuracy of the system and if
required a calibration (see
).
Note: It is the operator’s responsibility to keep a record of when and
what maintenance has been performed.
Note: If you suspect the BIOGAS 3000 system to have been damaged and
are unsure of the consequences of this, please contact our technical
support team at (800) 968-2026 or email
Note: Inspection of the catchpot and inline filter may be required more
frequently depending upon the application and the likelihood of liquid in
the sample gas.
Note: After performing a gas check or calibration, it may be necessary to
perform a further pressure test on the equipment to ensure it has
remained leak free.
Warning
Dependent upon the application, the equipment can
come in to contact with unsafe contaminants. It is
therefore recommended that suitable PPE is identified
and worn (such as gloves) and hands are washed
thoroughly after maintenance is completed.