The following steps describe how to replace an assembly in the network
analyzer.
Step 1.
Identify the faulty group. Begin with
Follow up
with the appropriate troubleshooting chapter that identifies the faulty
assembly.
Step 2.
Order a replacement assembly. Refer to this chapter.
Step 3.
Replace the faulty assembly and determine what adjustments are necessary.
Chapter 7, “Repair and Replacement Procedures.”
Step 4.
Perform the necessary adjustments. Refer to
Step 5.
Perform the necessary performance tests. Refer to
Step 6.
“Location of Calibration Stickers on PNA” on
to review where the calibration stickers should be placed on the
PNA.
Содержание N5221A
Страница 9: ...Contents Keysight N5221A 22A Service Guide 10 ...
Страница 10: ...11 Keysight N5221A 22A Service Guide Contents ...
Страница 132: ...5 8 Keysight N5221A 22A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 148: ...5 24 Keysight N5221A 22A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 6 4 Port Receiver Group ...
Страница 149: ...Keysight N5221A 22A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 7 2 Port Receiver Group ...
Страница 156: ...5 32 Keysight N5221A 22A Service Guide Theory of Operation Power Supply Group Operation 5 ...
Страница 264: ...6 108 Keysight N5221A 22A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 328: ...7 64 Keysight N5221A 22A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 346: ...8 18 Keysight N5221A 22A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...