4-24 Keysight N5221A/22A Service Guide
Troubleshooting
Rear Panel Troubleshooting
4-
d.
At the prompt, type
ping xxx.xxx.xxx.xxx
1
and press ENTRY
on the front panel. Refer to Step 5 for the results of a
successful ping.
5.
The analyzer attempts four cycles of communications with the indicated
LAN device.
—
It displays the time it took to complete each cycle.
—
Each cycle times-out after one second if no communication is
established and the message,
Request timed out
, is displayed.
—
It is common for the first of the four cycles to time-out even though
subsequent cycles pass.
—
See below for an example output of a successful ping.
C:>ping 141.121.69.162
Pinging 141.121.69.162 with 32 bytes of data:
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Reply from 141.121.69.162: bytes=32 time<10ms TTL=127
Ping statistics for 141.121.69.162:
Packets: Sent = 4, Received = 4, lost = 0 <0% loss>.
Approximate round trip times in milli-seconds:
Minimum = 0ms, Maximum = 0ms, Average = 0ms
6.
The above message verifies that one way communication from the
analyzer to the network has been established
7.
If the subnet mask was changed in step 2, set it back at this time.
How to Ping from the Local Area Network (LAN) to the Analyzer
Reverse communication should also be verified. Determining this, though, is
dependent upon your network setup and software. Generally, you need to
issue a ping command using the IP address of the analyzer to be tested. For
example, using Windows 95, 98, 2000, XP or 7 and while at a DOS prompt,
type in
ping xxx.xxx.xxx.xxx
2
. Then press ENTRY
on the front
panel. If full communication can be established, then the computer display
shows the cycle time for each of four cycle attempts (similar to that in step 5).
Other software may behave somewhat differently, but basically the same.
1. The letters x represent the IP address of the other device on the network.
2. The letters x represent the IP address of the analyzer.
Enter
Enter
Содержание N5221A
Страница 9: ...Contents Keysight N5221A 22A Service Guide 10 ...
Страница 10: ...11 Keysight N5221A 22A Service Guide Contents ...
Страница 132: ...5 8 Keysight N5221A 22A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 148: ...5 24 Keysight N5221A 22A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 6 4 Port Receiver Group ...
Страница 149: ...Keysight N5221A 22A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 7 2 Port Receiver Group ...
Страница 156: ...5 32 Keysight N5221A 22A Service Guide Theory of Operation Power Supply Group Operation 5 ...
Страница 264: ...6 108 Keysight N5221A 22A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 328: ...7 64 Keysight N5221A 22A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 346: ...8 18 Keysight N5221A 22A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...