7-50 Keysight N5221A/22A Service Guide
Repair and Replacement Procedures
Removing and Replacing the A51 Solid State Drive (SSD)
7-
Removing and Replacing the A51 Solid State Drive (SSD)
Certain unique files exist on the solid state drive (SSD) that are necessary for
proper operation of your analyzer. These files must be copied to another
location to allow them to be installed onto the new SSD after it has been
installed.
If you are replacing the SSD, the following procedure must be performed first.
Copy Unique Files from the Solid State Drive
If installing an SSD for Windows XP: if the user has loaded unique calibration
kit information, navigate to
C:\Program Files\Keysight\Network
Analyzer
and copy USER_CALKITFILE to a USB flash memory drive. Also copy
any personal user files that you wish to preserve.
If installing an SSD for Windows 7: for more information on the N8983A SSD,
refer to the
Windows 7 Operating System Upgrade Kit Installation Note
,
available online at
http://literature.cdn.keysight.com/litweb/pdf/N8983-90001.pdf
.
Tools Required
— T-10 TORX driver (set to 9 in-lb or 1.02 N.m; for all other T-10 applications)
— ESD grounding wrist strap
SSD Removal and Replacement Procedure
Refer to
for this procedure.
Removal
1.
Disconnect the analyzer power cord.
2.
Loosen the knurled nut (item
①
) and pull the A51 SSD out of the A17 CPU
board using the handle (item
➁
).
Replacement
1.
Using the handle (item
➁
), slide the A51 SSD into the A17 CPU board and
tighten the knurled nut (item
①
).
Figure 7-23
Solid State Drive Replacement
Содержание N5221A
Страница 9: ...Contents Keysight N5221A 22A Service Guide 10 ...
Страница 10: ...11 Keysight N5221A 22A Service Guide Contents ...
Страница 132: ...5 8 Keysight N5221A 22A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 148: ...5 24 Keysight N5221A 22A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 6 4 Port Receiver Group ...
Страница 149: ...Keysight N5221A 22A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 7 2 Port Receiver Group ...
Страница 156: ...5 32 Keysight N5221A 22A Service Guide Theory of Operation Power Supply Group Operation 5 ...
Страница 264: ...6 108 Keysight N5221A 22A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 328: ...7 64 Keysight N5221A 22A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 346: ...8 18 Keysight N5221A 22A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...