4-36 Keysight N5221A/22A Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
Checking the A11 13.5 GHz LO Synthesizer Output
1.
Refer to the block diagram at the end of this chapter and to
All Cables—All Options” on page 6-19
. Locate the flexible cable W41 at the
A11 LO synthesizer board.
2.
Disconnect W41 from J1207.
3.
Connect the spectrum analyzer to J1207.
4.
. Set the network analyzer for an 800 MHz CW
frequency and observe the spectrum analyzer measurement. For analyzers
with DSP version 4.0, an 807.61 MHz signal should be present. For
analyzers with DSP version 5.0, an 807.44 MHz signal should be present.
5.
If the observed problem was frequency banded rather than broadband
related, set the analyzer frequency to the center of the problem band. The
spectrum analyzer should measure a signal above the network analyzer
setting. For analyzers with DSP version 4.0, the signal is 7.61 MHz above
the network analyzer setting. For analyzers with DSP version 5.0, the
signal is 7.44 MHz above the network analyzer setting.
6.
If the LO signal is not present and the 50 MHz reference signal is present
from
“Checking the A10 50 MHz Reference Outputs,”
replace the A11 LO
“Removing and Replacing the A4–A13 Boards”
7.
If the signal is present, reconnect cable W41, and then continue with
“Checking the A21 HMA26.5 Output”
.
Checking the A21 HMA26.5 Output
1.
Refer to the block diagram at the end of this chapter and to
Cables, Standard 4-Port Configuration, Option 400” on page 6-56
the flexible cable W42 at the A21 HMA26.5.
2.
Disconnect W42 (4-port models) or W70 (2-port models) from the A21
HMA 26.5.
3.
Connect the spectrum analyzer to the open connector on the A21 HMA
26.5.
4.
. Set the network analyzer for a 800 MHz CW
frequency and observe the spectrum analyzer measurement. For analyzers
with DSP version 4.0, an 807.61 MHz signal should be present. For
analyzers with DSP version 5.0, an 807.44 MHz signal should be present.
5.
If the observed problem was frequency banded rather than broadband
related, set the analyzer frequency to the center of the problem band. The
spectrum analyzer should measure a signal above the network analyzer
setting. For analyzers with DSP version 4.0, the signal is 7.61 MHz above
the network analyzer setting. For analyzers with DSP version 5.0, the
signal is 7.44 MHz above the network analyzer setting.
Содержание N5221A
Страница 9: ...Contents Keysight N5221A 22A Service Guide 10 ...
Страница 10: ...11 Keysight N5221A 22A Service Guide Contents ...
Страница 132: ...5 8 Keysight N5221A 22A Service Guide Theory of Operation Synthesized Source Group Operation 5 Figure 5 3 Source Group ...
Страница 148: ...5 24 Keysight N5221A 22A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 6 4 Port Receiver Group ...
Страница 149: ...Keysight N5221A 22A Service Guide 5 25 Theory of Operation Receiver Group Operation Figure 5 7 2 Port Receiver Group ...
Страница 156: ...5 32 Keysight N5221A 22A Service Guide Theory of Operation Power Supply Group Operation 5 ...
Страница 264: ...6 108 Keysight N5221A 22A Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 328: ...7 64 Keysight N5221A 22A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 346: ...8 18 Keysight N5221A 22A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...