Error Messages
Error Messages
D.
Er
ro
r M
e
ssa
g
e
s
1201
CPU bd EEPROM write error
An error is generated while writing data to EEPROM.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
-100
Command error
A comprehensive syntax error has occurred for which the E4981A cannot detect further
details of the error. This error code simply indicates the occurrence of a command error
that is defined in IEEE488.2,11.5.1.1.4.
-140
Character data error
An error not included in the error numbers between -141 and -149 has occurred during the
syntax analysis of a character data element.
-148
Character data not allowed
Character data not allowed for this operation.
D
-104
Data type error
The parser has recognized impossible data elements. For example, numeric value or string
data is expected, but block data is sent.
-222
Data out of range
A data element (that does not violate the standard) has been received out of the range
defined for the E4981A.
-230
Data corrupt or stale
The data is invalid or a newly initiated read operation has not been completed since the
latest access.
E
-123
Exponent too large
The absolute value of the exponent has exceeded 32,000. (Refer to IEEE488.2,7.7.2.4.1.)
-170
Expression error
An error not included in the error numbers between -171 and -179 has occurred during the
syntax analysis of equation data.
-178
Expression data not allowed
An equation data element has been received where the E4981A does not accept any
equation data element.
-200
Execution error
A comprehensive execution error has occurred for which the E4981A cannot detect further
details. This error code simply indicates the occurrence of an execution error that is defined
in IEEE488.2,11.5.1.1.5.
Содержание E4981A
Страница 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Страница 2: ......
Страница 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Страница 14: ...12 Contents ...
Страница 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Страница 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Страница 96: ...94 Chapter6 Replacement Procedure 3 4 Rack Module Top Replacement Figure 6 3 Screw Fastening Sequence ...
Страница 98: ...96 Chapter6 Replacement Procedure PSU Deck Assembly Removale Figure 6 4 PSU Deck Assembly Removal ...
Страница 100: ...98 Chapter6 Replacement Procedure PSU Assembly Replacement Figure 6 5 PSU Assembly Replacement ...
Страница 102: ...100 Chapter6 Replacement Procedure Front Chassis Assembly Removal Figure 6 6 Front Chassis Assembly Removal ...
Страница 104: ...102 Chapter6 Replacement Procedure Front Panel Assembly Removal Figure 6 7 Front Panel Assembly Removal ...
Страница 108: ...106 Chapter6 Replacement Procedure Interface Board Assembly Replacement Figure 6 10 Screw Fastening Sequence ...
Страница 118: ...116 Chapter6 Replacement Procedure Power Inlet Assembly Replacement Figure 6 17 Power Inlet Assembly Replacement ...
Страница 122: ...120 Chapter6 Replacement Procedure Inverter Assembly Replacement Figure 6 20 Inverter Assembly Replacement ...
Страница 124: ...122 Chapter6 Replacement Procedure Display Support Removal Figure 6 21 Display Support Removal ...
Страница 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Страница 139: ...137 1 Chapter Title 2 Chapter Title 4 Chapter Title 5 Chapter Title C Power Requirement C Power Requirement ...
Страница 142: ...140 AppendixC Power Requirement Power Requirements ...