134
Appendix B
Firmware Update
Update the E4981A firmware using USB/GPIB interface
Update the E4981A firmware using USB/GPIB interface
Required Equipment
•
E4981A.
•
82357A/B USB/GPIB interface.
•
Personal Computer with I/O Libraries Suite (Windows 2000/XP).
How to update E4981A firmware using USB/GPIB interface
The following is the procedure to update the E4981A firmware using USB/GPIB interface.
Step 1.
Connect the USB/GPIB interface from the USB terminal in your computer to the GPIB
terminal in the E4981A rear panel. Then turn the E4981A on.
Step 2.
Copy the following file from Keysight Technologies web site to the temporary work folder
in your computer’s HDD.
•
fw_upd.exe
Step 3.
Execute the
Command Prompt
on your computer.
Step 4.
Change the current folder to the temporary working folder with the “fw_upd.exe” file
using
cd
command.
Step 5.
Enter
fw_upd.exe GPIB0::17::INSTR
command to execute the firmware update program.
The numerical value of “17” in the command is GPIB address of the E4981A. Then, press
the Enter key.
Step 6.
Wait a few minutes until the command prompt is displayed again.
Step 7.
Press
System
key on the E4981A front panel to check the firmware revision.
Содержание E4981A
Страница 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Страница 2: ......
Страница 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Страница 14: ...12 Contents ...
Страница 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Страница 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Страница 96: ...94 Chapter6 Replacement Procedure 3 4 Rack Module Top Replacement Figure 6 3 Screw Fastening Sequence ...
Страница 98: ...96 Chapter6 Replacement Procedure PSU Deck Assembly Removale Figure 6 4 PSU Deck Assembly Removal ...
Страница 100: ...98 Chapter6 Replacement Procedure PSU Assembly Replacement Figure 6 5 PSU Assembly Replacement ...
Страница 102: ...100 Chapter6 Replacement Procedure Front Chassis Assembly Removal Figure 6 6 Front Chassis Assembly Removal ...
Страница 104: ...102 Chapter6 Replacement Procedure Front Panel Assembly Removal Figure 6 7 Front Panel Assembly Removal ...
Страница 108: ...106 Chapter6 Replacement Procedure Interface Board Assembly Replacement Figure 6 10 Screw Fastening Sequence ...
Страница 118: ...116 Chapter6 Replacement Procedure Power Inlet Assembly Replacement Figure 6 17 Power Inlet Assembly Replacement ...
Страница 122: ...120 Chapter6 Replacement Procedure Inverter Assembly Replacement Figure 6 20 Inverter Assembly Replacement ...
Страница 124: ...122 Chapter6 Replacement Procedure Display Support Removal Figure 6 21 Display Support Removal ...
Страница 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Страница 139: ...137 1 Chapter Title 2 Chapter Title 4 Chapter Title 5 Chapter Title C Power Requirement C Power Requirement ...
Страница 142: ...140 AppendixC Power Requirement Power Requirements ...